Instruction Manual of PortableOxygen Concentrator
Model: KPC Oxygen Concentrator
Please read the instruction manual carefully before using the product.
WebSite: https://dedakjoxygenconcentrators.com/
Mail Address: dedakj@yahoo.com
*Please confirm that the instructions have been carefully read and fully understood before using this product.Note: When using this product, no smoking, no use in a fire The product built-in humidifier bottle is strictly prohibitedto join the flow of pure water, can increase oxygen humidity by lockingwater cotton.
- The expected service life of the product is 5 years. The service life depends onthe service environment and later maintenance, Poor use environment willshorten the service life of the equipment.
- Some users cannot hear or see alerts or communicate discomfort, so additionalmonitoring or attention is required by other guardians when using this device.
- This product needs oxygen inhalation by connecting the appropriate nasal oxygen tube. Each equipment is equipped with a disposable nasal oxygen tubeAn anti-extrusion nasal oxygen tube suitable for this machine interface can beselected as required, which must be a registered medical device. increasing thelength of the tube can reduce the noise in the process of oxygen transmission;when the nasal oxygen tube is used for lengthening, it may be necessary toincrease the set flow level. Please follow the doctor's advice.
- Please follow the instructions for the correct use of nasal oxygen tube.
- Due to the different materials used in nasal oxygen tubes, some nasal oxygentubes may have odor.
- This product cannot be connected to other devices that are not allowed by themanufacturer.
- Itis forbidden to use power sources or power lines, power adapters or otheraccessories outside the specified, and the use of non-specified accessories maycause security risks or damage equipment performance.
- Unless specified, all accessories of this product must be provided by thedesignated seller.
- Ifthe accessories are damaged or lost, please contact the equipment provider,
- Please use this product according to the expected use required in theinstructions.
- In some cases, excessive inhalation of oxygen is dangerous. Please follow the doctor's advice when using it.
- Feel uncomfortable in the use of this product, please immediately stop usingand medical treatment.
- Oxygen helps to burn, smoke and fire in the presence of strictly prohibited useof this equipment.
- This product cannot be used for life support and life extension, and is notsuitable for newborns and infants.
- Please do not expose the equipment to rain or snow. Please do not operate theoxygen generator in the rain, which may lead to electric shock and equipmentdamage.
- Please do not use this product in high temperature and high humidityenvironments (such as unmanned cars in high temperature environments orhigh humidity bathrooms) to avoid damaging equipment.
- Non-professionals should not disassemble the oxygen generator at will. Anychanges to the device can damage performance or damage the device and willinvalidate your warranty.
Warning: This product should not be close to or stacked with otherequipment. if it must be close to or stacked, it should beobserved and verified that it can operate normally under theconfiguration of its use.
Warning: In addition to the cables sold by manufacturers as spare partsof internal components, the use of accessories and cablesother than the provisions may lead to an increase inequipment emission or a decrease in immunity.
Warning: Patients with strict requirements for oxygen concentrationsshould follow the guidance of professional doctors andmonitor the alarm instructions of equipment.
Warning: Patients with severe lung disease, choose what oxygenuptake should consult a professional physician.
(1). Maintenance
- The maintenance cycle of portable oxygen generators is approximately one yearand only professionals in maintenance centers, such as authorized personnel orstaff trained in factories, can be repaired or debugged.
(2).Radio frequency interference
- Most electrical appliances are vulnerable to radio frequency interference, so theuse of portable communication equipment near the oxygen generator maycause interference with the machine.
(3).Personal injury risks such as burning, electric shock andfire
In order to reduce the risk of personal injury, please note the following:
- Do not use in shower.If the patient needs continuous use, the oxygen machinemust be placed at least 2.0 M away from the bathroom.
- Don't touch the oxygen maker when your body is wet. Do not use or storeoxygen generators near easily falling water or other conductive liquids.
- It is forbidden to contact the oxygen generator that falls into the water or otherconductive liquids. Ifit falls into it, please immediately remove the adapterpower plug.
- When the machine is not used, the power plug must be removed. When thedevice is not used for a long time, the battery needs to be charged 70 %. Thebattery and power plug must be removed, and the oxygen generator, batteryand adapter should be placed in the dry place.
- Guardianship must be used when there are children or persons in need ofmobility.
- Please use this machine according to the expected use of the products in theuser manual.
(4).Waste Consumables Branch
- When the relevant waste such as the breathing pipe and filter used in theoxygen generator is used, please implement the product in accordance with therelevant regulations of the local government to have no impact onenvironmental protection and pollution. Forelectronic and electronic waste such as batteries, circuit boards, etc, ) do not discard at will, please contact therelevant departments in accordance with the relevant provisions of the localgovernment, and properly handle according to the corresponding instructions.
1.Safety requirements
2.Safety requirements
- (1)Maintenance.
- (2)Radio frequency interference
- (3)Personal injury risks such as burning, electric shock and fire
- (4)Waste Consumables Branch..
3.Product introduction
4.Product principle
5.Product structure
6.Product working conditions
7.Power supply parameters
8.Product transportation and storage conditions
9.Reserve oxygen sources
10.Product performance
11.Scope of use /intended use, contraindications
12.Display and alarm functions
13,Instructions for installation and use
14.Operational guidance
15.Operational guidance
- (1).Connecting power supply
- (2).Start-up preheating.
- (3).Connecting nasal oxygen tube
- (4).Adjusting the flow level
- (5).Shutdown
16.Maintenance and cleaning
- (1)Maintenance and cleaning of structural components
- (2)Battery replacement and maintenance
- (3).Replacement of nasal oxygen tube
- (4).Term of use
- (5)Maintenancecycle
17.Detailed parameters of equipment
18.Common problems and solutions
19.Packing list
20.Box removal note
21.Notes on warranty
- Compact portable flow oxygen machine.
- Real time concentration display, timing function.
- 3.5 inch touch screen, menu management
- Unique mute Technology
- Portable oxygen generator refers to the equipment that uses molecular sievepressure swing adsorption principle to improve oxygen concentration byadsorbing nitrogen and other gas components. When the equipment workscompressed air is injected into a closed adsorption tower equipped withmolecular sieve, resulting in an increase in the pressure in theadsorption tower.The molecular sieve adsorbs a large amount of nitrogen in the compressed airwith the increase of environmental pressure, while the oxygen in thecompressed air still exists in the form of gas and is collected through a certainpipeline. This process is usually called “adsorption" process. When theadsorption of nitrogen by molecular sieve in the container reaches the critical state of adsorption saturation, the adsorption tower is reduced by blowingpressure. With the decrease of environmental pressure, the ability of molecularsieve to adsorb nitrogen decreases, and nitrogen is released from the molecular sieve as exhaust gas, This process is usually called desorption. In order to ensurethe continuous and stable output of oxygen, the oxygen generator usually usestwo (or more) molecular sieve adsorption towers. Through the control of rotaryseparation valve, one adsorption tower is in the adsorption process while theother is in the desorption process. The two work alternately to complete thecontinuous oxygen production process.
- Portable oxygen generator consists of compressor, battery, solenoid valve.molecular sieve, circuit control system, heat dissipation device, flow control device, oxygen delivery mask /nasal oxygen tube (purchased medicalequipment).
- Working temperature range : -10 ~ 40 °C
- Working humidity range : ≤ 80 %
- Working pressure range : 86 kPa ~ 106 kPa
- AC power supply:AC.100-240 V,AC 50/60 Hz ;
- Internal battery : DC 14.8V ± 10 %
- Transport and storage temperature range : -20 ~ 55 °C
- Transport and storage humidity range : 5 %-90 % non-condensable statetransport in dryenvironment
- Storage pressure range :670hPa~ 1060hPa
- It is recommended that you provide spare oxygen sources for equipment toprevent mechanical failure or battery depletion.
- Scope of application/expected use:
Oxygen-enriched air was produced by molecular sieve pressure swingadsorption process with air as raw material. The oxygen concentration rangewas ≥ 90 % (V/V), providing oxygen therapy or alleviating discomfort caused byhypoxia.
Warning: professional doctor guidance should be followed when
Warning: Do not use this device when flammable anesthetics exist
- Contraindications :
Contraindications: oxygen poisoning, oxygen allergy patients prohibited
Warning: Professional doctor guidance should be followed when using.
- LCD screen interactive main page
After startup, the system enters the main page, and the left column is the menu,which is divided into settings, modes, flow rate, timing and analyse.The top of the page is oxygen concentration, flow display and power display.Main screen display concentration settings, gear settings.
- Setup page
Click the LCD screen Set up button to enter the device information page, deviceinformation page display as follows:
Alarm sound size setting, set machine alarm sound;Pulse sensitivity settings, according to the use of environment and personalhabits, you can set the respiratory pulse sensitivity.Machine brightness adjustment.
- Mode page
Click the mode button of the LCD screen to enter the device information page.
The device information page is displayed as follows:
Pulse /fixed frequency settings, set to pulse mode, the device detects thebreathing pulse, when detected inspiratory instantaneous release of oxygen,when breathing oxygen storage; after setting to a constant frequency mode, theequipment emits oxygen once in two seconds, and the breathing pulse is nolonger detected, which is suitable for wearing masks, strong wind environmentand the environment requiring continuous oxygen output.
Plateau mode setting, in the plateau environment, can open the plateau mode, provide sufficient oxygen.
- Timing Page
Click the timing button of the LCD screen to enter the device information page.
The device information page is displayed as follows:
Set the time in hours and minutes to enable the open. The setting range ofhours is 0~ 60; the setting range of minutes is 1 ~ 59.Set to complete click the key, return to the main page;
- Analysis page
Click the LCD screen analysis button to enter the device information pagedevice information page display as follows:
Machine temperature: equipment machine internal temperature;
Battery temperature: batteryinternal temperature;
Respiratory frequency: respiratory frequency detected by equipment;
The running time: the running time of the equipment from the start to now;
Total running time: how long did the equipment run from the factory;
Hardware version: Hardware internal version number;
Software version: System software version number;
Click Return to return to the main page.
Alerting Message: (Interpretation and Operation ofAlarm Content)
Fan shutdown: The fan stops turning, the fan stopsturning will cause the machinetemperature to rise rapidly, pleaseshut down in time and contact themanufacturer.
Hyperpyrexia: If the temperature is too high alarmplease check whether theenvironment temperature is too high, turn off 10 to 20 minutes later, start again, if the temperature is still toohigh alarm, please contact themanufacturer in time.
Compressor shutdown: Compressor not started, pleasecontact the manufacturer in time
Low oxygen concentration: Please contact the manufacturer intime
Low battery voltage: Please charge in time
- Please remove all packaging for initial use and place plastic bags where childrencannot reach them so as not to cause asphyxiation.
- Be sure to read the instructions carefully before use to ensure that the parts areinstalled in place and correctly operated and used.
- The product in the use of humidified bottles to humidify oxygen is strictlyprohibited to join the flow of liquid water, only wet locking medical cotton intothe leak net to humidify the oxygen in the humidified bottle.
- When using this product, please do not touch any live components such asplugs, power lines, adapter connections and other live parts with wet hands, inorder to avoid causing electric shock damage.
- Please do not damage or break the power line or plug, and do not use plugs withloose ends to avoid fire or electric shock.
- Itis forbidden to place heavy objects on this equipment so as not to bedamaged.
- Please do not place any part of the body near the exhaust port or other feverparts for a long time, so as not to cause burns.
- Please set the appropriate flow file according to the doctor's advice.
- When the product is used, the inlet should be located in the good ventilation,and the inlet and exhaust ports cannot be blocked, so as to avoid heataccumulation affecting product performance and even causing fire.
- This product should avoid direct sunlight, so as not to cause excessive localtemperature.
- This product should be used away from the fire, no smoking, so as not to causefire.
- Long-term operation in a humid environment may shorten the life of molecularsieves.
- This product should be away from pollution, smoke and flammable, explosivevolatile goods, such as alcohol, gasoline and other dangerous goods, in order toavoid fire or explosion.
- This product cannot be side-laid or inverted when used
- The product is used for 5 years from the date of production. Consumable lifeand use conditions related, please according to the actual use of timely purchaseand replacement of consumables, production date see product growth qualifiedcard.
- Failure to use the equipment in the prescribed manner may cause damage tothe equipment and render warranty ineffective.
- Part name
A : LCD screen B : start button C : oxygen outlet D : inlet style gate E : battery box F : battery box buckle G : power adapter interface H : wet bottle side cover l : wetbottle window J : outlet
Note: Exhaust hot air is a normal phenomenon. Itis forbidden to block theinlet and outlet.
The steps of removing wetting bottle
Pull the side cover along the arrow
1, Pull out PE wet bottle cap.
2, Take out PVA medical lockwater cotton in order, leakage.
3, 3. Put the lock cotton in purewater to absorb water fully.
4, First, 50 % of the pure water into the water-locked cotton wasextruded and then put into theleakage net (note: non-flowablewater out-flowedfrom the bottomof the leakage net). Then, theleakage net was put into thewetting bottle. Finally, the PEcover was pressed to ensure thesealing of the wetting bottle.
Note: Installation steps of wetted bottles arerecounted strictly according to disassembly steps.
- Disassembly Method of Battery Components
1.Press the button in the first step in the direction of the arrow(1).
2.The second step is to hold the battery to pull out the battery component in the direction of arrow (2) while holding the button.
- Disassembly Method of Drying Components
Pull the side cover along the arrow
Drying box
1, Open the lid as shown in Figure1.
2, As shown in Figure ,remove the cover, seal silica gel ring, drying box, filter cotton.
3, Cleaning filter cotton (fully controlling drymoisture).
4, The steps of installing the drying box arestrictly followed by the steps of splitting.
(1). Connecting power supply
- Select the appropriate power connection conditions according to the use environment
A,When use battery only
Install the special battery to the battery interface, Please note that the battery should be installed firmly.
B,When using AC power adapter
1, Connect the AC power adapter to the power line firmly and connect the input plug to the socket. When the power indicator light on the adapter is on, it indicates that the power connection is normal.
2, Requirements for input: AC voltage 100-240V, current 2.0-1.0A, frequency 50/60Hz.
3, The output plug of AC power adapter is connected to the DC power input interface of the equipment.
Note: When you don't use batteries, pay attention to protecting the battery interface and the power on the device
Pool interface, please do not contact the conductor or directly touch with hand.
Note: When battery power is used alone, the equipment has been battery capacity empty' after the automatic shutdown, do not use this air battery alone to restart operation, so as not to affect battery life and equipment performance.
Warning: Prohibit the use of non-product lithium batteries.
(2).Start-up preheating
- Press the device to turn on the key, heard a ' whisper ' of the boot prompt, into the boot screen, system startup equipment into preheating. Thepreheating time of the equipment is about 2 min. When the equipment statereaches the normal use requirement after the preheating, it can be worn wellwith nasal oxygen tube for normal use.
Note: During the machine preheating, the equipment willautomatically spray oxygen to empty the internal air, whichWhen the device output flow and concentration may not meetthe standard, please do not connect to the use of nasal oxygentube.
(3).Connecting nasal oxygen tube
- The horn end of the nasal oxygen tube equipped in the factory is connected tothe metal outlet nozzle of the equipment to ensure that the connection isreliable and does not leak. Attention is paid to the nasal oxygen tube not to twistor block, so as to avoid causing equipment-related alarm and affecting normal use.
- If you need to buy the nasal oxygen tube by yourself, please contact theequipment provider or under the guidance of professional medical staff, selectthe nasal oxygen tube suitable for the equipment and obtain the formal medical registration certificate.
Note: This equipment is pulse oxygen supply mode, using thisequipment, must connect nasal oxygen tube use.
Note: In order for the oxygen generator to correctly detect breathingand deliver pulse oxygen, please do the nasal oxygen tube is properlyinstalled to ensure that the oxygen tube is not twisted or blocked.
Note: Please follow the manufacturer instructions of nasal oxygentube. Supply bymanufacturer or equipment that it is recommendedto replace the nasal oxygen pipe and other accessories can bepurchased from the equipment supplier.
Note: Do not use unfit equipment or no equipment provider or not inprofessional care purchase of nasal oxygen tubes under the guidanceof personnel to avoid affecting the normal operation of equipmentand user normal use.
(4).Adjusting the flow level
- Please set the flow level according to the oxygen flow recommended by thedoctor.
- Before the doctor makes suggestions, please communicate with the doctor theperformance of the device, so as to avoid the doctor put forward suggestionsthat are not suitable for the device.
(5).Shutdown
- When the device works normally, press the switch key for about 3 seconds, thedevice sends a'ti 'prompt and enters the shutdown interface.
(1)Maintenance and cleaning of structural components
Note: In the maintenance and cleaning work must first shut down the equipmentto cut off the power supply and take off the battery.
- The dust at the inlet grille of the equipment is cleaned regularly by brush orvacuum cleaner.
- The wetting bottle must be cleaned and cooled after each use, and the water.locked cotton in the wetting bottle must be cleaned and cooled after each use(Note: Water lock cotton cleaning clean can be reused, recommended not morethan 10 times.)
- Air filter cotton is cleaned every half month, Note that a large amount of dustmay be absorbed and accumulated by the intake filter cotton after long-termuse. Please be careful to operate and do a good job of protection to avoidinhalation of dust. Please properly handle abandoned filter cotton.
- Failure to install or replace filter cotton for a long time may affect equipmentperformance or damage equipment
- Filter cotton, lock water cotton, drying agent as consumables, need to buytimely contact with equipment suppliers. When the equipment prompts theneed to replace the molecular sieve, please contact the designated seller intime.
- Permanent placement may shorten the lifetime of molecular sieves.
- Long-term operation in humid environment may shorten the lifetime ofmolecular sieve.
- When the molecular sieve is close to the service life, it may cause increasedinternal pressure, increased noise and decreased oxygen concentration,affecting the equipment performance and normal use. Please payattention tothe equipment noise, oxygen concentration and other related information intime.
Warning: Please do not change molecular sieve by private disassembly. Onlyequipment suppliers or qualified maintenance engineers can change molecularsieve by private disassembly.
Note: Molecular sieves are consumables. Please contact the equipment provider intime for replacement.
- Users can purchase nasal oxygen tubes on their own, as required, but ensurethat the following conditions are fully met :
- 1.Please purchase from the formal channels and obtain the registration certificate of medical devices for nasal oxygen tubes;
- 2.The gas flow rate in the nasal oxygen tube needs to reach 5 liters per minuteto adapt to the product. Please wear the nasal oxygen tube correctly and use itaccording to the instructions of the nasal oxygen tube.
(2)Battery replacement and maintenance
- This equipment needs to use a special battery equipped in the factory, and abattery is equipped in the factory, Users can contact the designated seller topurchase batteries as required.
- Battery life: 1 year.
- When not used for a long time, please remove the battery from the equipmentand pay attention to the protection of the battery electrode, not contact withthe metal and other conductors, to prevent the risk of fire.
- When the battery is stored, please put it where children cannot reach to avoiddanger.
- Long-term unused batteries need regular charging and maintenance, and it issuggested that the battery capacity should be maintained at about 70 %.
Warning: Prohibit the use of non-product lithium battery
Note: The battery is a consumable material. Please contact theequipment provider in time when it needs to be replaced.
(3) Replacement of nasal oxygen tube
- The machine is equipped with nasal oxygen tube. The nasal oxygen tubeequipped in the factory should have one of the following medical deviceregistration certificates :
- Users can purchase nasal oxygen tubes on their own, as required, but ensurethat the following conditions are fully met :
1, Please contact equipment providers or purchase under the guidance of professional medical staff.
2, Please purchase from regular channels and the nasal oxygen tube has obtained amedical device registration certificate ;3,The rated flow rate of oxygen in the nasal oxygen tube is 5 liters per minute.
(4) Term of use
Main oxygen generator: 1 years
Molecular Sieve: 1 year
Battery: 1 year
Nasal oxygen tube: See product packaging
When the oxygen generator is close to the service life, it may lead to equipment performance degradation or equipment failure. Please pay attention to flow,concentration, fault and other related alarm information.
When the molecular sieve is close to the service life, it may cause the increase of internal pressure and the decrease of oxygen concentration. Please pay attention to the pressure, concentration and other related alarm information in time.
When the battery is close to the service life, it may lead to abnormal conditions such as unable to charge, slow charging, unable to discharge, and sharp decline in endurance time. Please pay attention to the battery state and other related information in time.
Note: The use period in the table is the recommended value, and the actual use period will follow the actual
Use environment and actual usage changes, please pay close attention to equipment usage and related alerts.
(5) Maintenance cycle
Name | Maintenance Period | Maintenance Mode | Remark |
Whole intake grille | Weekly | User cleaning | Note: cleaning with the brush or vacuum cleaner |
Air inlet filter cotton | Half a month | Buy new one from the official store | There is a total of 1 piece out of the factory |
Molecular | Once a year | Buy new one from the official store | |
Nasal oxygen tube | |||
Battery | Buy extra one from the official store | No special maintenance for normal use |
Note: The maintenance cycle in the table is the recommended value, and theactual maintenance cycle changes with the actual use environment and actual use,please pay close attention to the equipment use status and related alarm prompts.
Product Name | portable oxygen concentrator |
Product Accumulation Size | L*w*h:178mm*82mm*211mm |
Product weight | 2.35Kg ± 0.1Kg (including a battery) |
User Page | 3.5 -inch color touch LCD screen |
Running Noise | ≦ 60dB(A)(gear 8 ) |
Warm-up Time | 3 minutes |
Oxygen Concentration | 93%±3% |
Respiratory Rate | 10-40 BPM |
Maximum Export Pressure | 140KPa+20% |
Alternating Current Power Supply | AC100-240V 50/60 Hz 100V A |
Internal Battery | 6700 mAh, 14.4VDC 8A Max |
Battery Duration | Up to 3.5 hours ( 1 gear to run ) |
Battery Charging Time | Up to 3.5 hours |
Suitable Working Environment ForEquipment | Suitable temperature :5~35 °C;Suitable humidity : 10 %~60 %, non-condensable atmospheric pressurerange :670hPa~1060hPa |
Suitable Environment For NormalStorage Of Equipment | Suitable temperature : 5 ~ 40°C;Suitable humidity :10 %~90 %, non-condensable atmospheric pressurerange :670hPa~1060hPa |
Transportation Request | Keep dry, be careful, don 't reverse orsideways, don 't roll |
Air Environment Requirements | Equipment that cannot be used in thecase of flammable anesthetic gas mixedwith air or flammable anesthetic gasmixed with oxygen or nitrous oxide |
Question | Possible Reasons | Recommended Solutions |
The device cannot open | Battery not installedcorrectly | Remove the battery andinstall it correctly |
Battery depletion | Insert battery, connectadapter to charge battery | |
Poor AC powerconnection | Check power connection; Check if the green light ison | |
Equipment failure | If the above operationcannot solve the problem, please contact yourequipment supplier | |
No oxygen output | Equipment not opened | Open the oxygengenerator |
Nasal oxygen tube knotor obstruction | Check the nasal oxygentube and its connectionto the oxygen outlet | |
Wetted bottle notinstalled | Correct installation ofwetting bottle | |
Equipment failure | Contact equipmentsuppliers | |
Insufficient oxygenconcentration | The equipment ispreheating | Wait for 3 minutes; if thisproblem has not beensolved, please contact theequipment supplier. |
Molecular sieves mayneed repair | Please contact equipmentsupplier to replacemolecular sieve bed. |
Note: the following state is non-fault:
The exhaust port is the heat dissipation port. When the equipment runs for a long time or the ambient temperature is high, thetemperature of the gas discharged by the equipment will increase, which is anormal phenomenon, The machine has high temperature protection. if the exhaustport is blocked or other reasons lead to excessive temperature of theequipment,the equipment will alarm and automatically shut down.
Serial Number | Name | Quantity |
1 | Main oxygengenerator | 1 |
2 | Rechargeable LithiumBattery | 1 |
3 | Carrying backpack | 1 |
4 | AC adapter(power lineattached) | 1 |
5 | Battery Power Cord | 1 |
6 | Car charger | 1 |
7 | Nasal oxygen tube | 1 |
8 | Air inlet filtercotton | 1 |
- Check whether the packing box is damaged or not, and inform shippingcompany and equipment supplier in time if damaged.
- Carefully remove the machine and related parts, and compared with the packinglist.Ifthere are parts inconsistent with the packing list or quality problems,please contact the equipment provider or after-sales service.
- Please keep packing boxes and packing accessories forstorage andtransportation.
- Molecular sieves, batteries and AC adapters are guaranteed for 1 year from thedate ofpurchase.
- During the warranty period, non-human causes of failure, free warranty.
- Reimbursable maintenance services are provided if maintenance costs arecharged outside the warranty period and damage is caused by improper use. When the product needs warranty, please send the machine and the warrantycard back to the manufacturer, the transportation cost is borne by the user.